Model CPB5800

Dead-weight tester (SEO-Seite)

Hydraulic version to 1,400 bar [20,000 lb/in²]

Selected documents
Sales contact Questions? Find a contact in your vicinity.

Applications of the dead weight tester

  • Primary standard for calibrating the pressure scale in a hydraulic range up to 1,400 bar [20,000 lb/in²]
  • Reference instrument for factory and calibration laboratories for the testing, adjustment and calibration of pressure measuring instruments
  • Complete, stand-alone system, also suitable for on-site use

Special Features of the dead weight tester

  • Total measurement uncertainty to 0.006 % of reading
  • Extremely flexible instrument with a wide range of single and dual-range piston-cylinder systems
  • Dual-range piston-cylinder systems with fully automated changing between ranges
  • Factory calibration includes traceability to national standards, as standard, with UKAS calibration possible as an option
  • Fast and safe replacement of the piston-cylinder system via patented ConTect quick-release system as an option

Description of the dead weight tester

Proven primary standard
Dead weight testers are the most accurate instruments available on the market for the calibration of electronic or mechanical pressure measuring instruments. The direct measurement of the pressure (p = F/A), as well as the use of high-quality materials enable a very small measurement uncertainty, in conjunction with an excellent long-term stability.

It has therefore been used for years in factory and calibration laboratories in industry, national institutes and research laboratories.

Stand-alone operation
Due to its integrated pressure generation and the pure mechanical measuring principle, the model CPB5800 dead weight tester is ideal for on-site use for maintenance and service.

Basic principle
Pressure is defined as the quotient of force and area. The core component of the model CPB5800 dead weight tester is therefore a very precisely-manufactured piston-cylinder system, which is loaded with masses in order to generate the individual test points.

The mass load is proportional to the target pressure and this is achieved through optimally graduated masses. As standard, these masses are manufactured to the standard gravity (9.80665 m/s²), though they can be adjusted to a specific location and calibrated according to UKAS accreditation.

DH-Budenberg
Dead weight tester CPB5800

Dead weight tester CPB5800

  • Dead weight tester CPB5800
Wish list - Documents (0)

There are currently no documents on your wish list. You can add any document from the download area to your wish list. To do this, select the desired language using the flag symbol.

The maximum number of 20 entries was reached.

To receive your download link via e-mail, enter your address here:

The e-mail address is invalid.